This job aid provides instructions to add comments to QC measurements and edit incomplete QC measurements from the POCcelerator™ Web QC Monitor screen.

POCcelerator™ Web – SIEMENS Healthineers Add Comments and Edit Incomplete QC Measurements Job Aid Add Comments to QC Measurements QC monitor 1. Select QC > QC monitor 2. Use the simple and/or advanced filter settings to Time range: Today 2 day 7 days 1 month From: 1/1/2022 To: 4/11/2022 filter the QC measurements for selection Ward: Emergency; ER Room 1 2 Sample type: ---- 3. Select the Details button for the chosen QC Model name: ---- QC level: Device ID: - -- - QC lot: measurement Analyte: ---- 4. Select a Lab comment from the dropdown list or Print by device Print by model All QC measurements Incomplete QC measurements enter a lab comment manually All measurements 5. If a QC violation is to be released, check the Measurement time Location Model name Device ID QC level C lot Analyte Target Range by Range by manufacturer law QC result value Unit User Released checkbox 2/18/2022 7:06:05 AM ER Room 1 (Emergency) RapidPoint 5xx 1001 Level 3 AQC1 Ca++ 0.80 0.70 - 0.90 ----- 0.80 mmol/L AQC, AQC (AQC) Details 3 • By checking the Released checkbox, measurement methods are not blocked despite 10/24/2018 ER Room 1 RapidPoint Level 12:56:44 PM (Emergency) 5xx 1001 2 AQC1 Ca++ 1.20 1.10 - 1.30 1.11 - 1.29 1.20 a QC violation, and patient measurements can mmol/L AQC, AQC (AQC Details Details: be performed before a valid quality Ward: ER Room 1 (Emergency) Sample type: Whole blood Transmission time: 10/24/2018 12:56:49 PM measurement result is provided Model name: RapidPoint 5xx Level: Level 2 User. AQC, AQC (AQC) Device ID: 1001 QC lot number: AQC1 User comment: SYSTEM 6. Confirm the input by selecting the green Device: RapidPoint 500;1001 Reagent lot number: 654321 4 Lab comment: checkmark button Released: By user: 6 X 5 1 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 POCcelerator™ Web – SIEMENS Healthineers Add Comments and Edit Incomplete QC Measurements Job Aid Edit Incomplete QC Measurements All QC measurements Incomplete QC measurements 2 Measurement time Device name Ward Analyte Target Range by Range by manufacturer law QC result value Unit 1. Select QC > QC monitor 6/12/2011 10:22:41 AM ABL8D;300587 General Station (Department) Citrate APTT 31.21 3 Modify 2. Select the Incomplete QC measurements tab 5/22/2011 12:01:02 PM ACCU-CHEK Inform II:UU11002445 General Station (Department) Citrate PT 30.00 10 - 50 ----- 35.24 S Modify 3. Select the Modify button for the chosen QC 6/30/2010 2:26:38 PM ACCU-CHEK Inform II;UU11002445 General Station (Department) Citrate PT 30.00 0 - 6 ----- 33.08 Modify Details: Sub lot number is missing! measurement Data is locked! Ward: General Station Sample type: Vollblut Transmission time: 7/9/2010 2:26:39 PM 4. Edit the QC measurement (Department) Model name: ACCU-CHEK Inform II Level: Level 1 User: Tec Documentation [1000) 5. Confirm the input by selecting the green checkmark Device ID: UU11002445 Reagent lot number: 320057 User comment unknown error - repeat measurement button Device: ACCU-CHEK Inform QC Solution lot II;UU11002445 number: --- - X 5 4 2 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022 SIEMENS Healthineers Please note that the learning material is for training purposes only. For the proper use of the software or hardware, please always use the Operator Manual or Instructions for Use (hereinafter collectively “Operator Manual”) issued by Siemens Healthineers. This material is to be used as training material only and shall by no means substitute the Operator Manual. Any material used in this training will not be updated on a regular basis and does not necessarily reflect the latest version of the software and hardware available at the time of the training. The Operator Manual shall be used as your main reference, in particular for relevant safety information like warnings and cautions. Please note: Some functions shown in this material are optional and might not be part of your system. Certain products, product related claims or functionalities (hereinafter collectively “Functionality”) may not (yet) be commercially available in your country. Due to regulatory requirements, the future availability of said Functionalities in any specific country is not guaranteed. Please contact your local Siemens Healthineers sales representative for the most current information. The reproduction, transmission or distribution of this training or its contents is not permitted without express written authority. Offenders will be liable for damages. All names and data of patients, parameters and configuration dependent designations are fictional and examples only. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. POCcelerator and all associated marks are trademarks of Siemens Healthcare Diagnostics Inc., or its affiliates. All other trademarks are the property of their respective owners. © Siemens Healthcare GmbH 2022 Siemens Healthineers Headquarters Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen, Germany Telephone: +49 9131 84-0 siemens-healthineers.com 3 Effective Date: 20 Apr 2022 | HOOD05162003270754 Unrestricted © Siemens Healthineers, 2022

  • poccelerator
  • poccelerator web
  • pocweb
  • qc monitor
  • qc